Relative Humidity Calibration Kit
Factory
Calibration:Each Kestrel Pocket weather Meter is individually calibrated after assembly. In essence, each unit is "taught" what the correct relative humidity (RH) values should be for given signals from the RH sensor. NK calibrates at RH=32.8% and RH=75.3%. Calibration is done in two sealed environmental chambers maintained at 25°C. The humidity is held constant with saturated aqueous salt baths: magnesium chloride for the RH 32.8% chamber and sodium chloride for the 75.3% chamber. Chamber temperatures are held constant at 25°C ± 0.5°C The temperature and RH of each chamber is independently monitored using a precision chilled mirror dew point hygrometer (EdgeTech Model 2002) accurate to ± 0.1Cº and NIST (National Institute of Standards and Technology) traceable. A proprietary auto calibration routine controls the calibration process. After 60 minutes this routine stores a calibration value which forces the Kestrel to be correct at the RH value of the chamber. The response of the
humidity sensor is a highly reproducible, non-linear curve. This non-linear
characteristic is stored in the microprocessor used to compute RH, and in
this manner we are able to maintain an accuracy of ± 3% of scale over the
relative range 5% - 95%.
Field Calibration:
Please read the entire instructions before attempting to calibrate your Kestrel. |
||
Contents of the Humidity Calibration
Kit: The large container is used to create a stable environment. The small containers are used to hold the saturated salt solution. The saturated salt solutions generate well-defined humidities when they are confined to a contained environment. |
![]() |
|
|
Setting up the Humidity Chambers:
Setting up the Kestrel 3000: Before recalibrating the Kestrel, the expected humidity values must be programmed into the unit:
Running the Calibration Routine: Once the expected RH values have been programmed, the unit may be recalibrated:
If the ON button is pressed at any time during the above routine, the unit will shut off and the re-calibration will be aborted.
Setting up the Kestrel 4000: Before recalibrating the Kestrel, the expected humidity values and calibration time must be programmed into the unit:
Running the Calibration Routine for the Kestrel 4000
Your Kestrel 4000 is now recalibrated and ready to take accurate measurements. If the unit is
turned off at any time during the above routine, the re-calibration will be
aborted.
RH Verification:
It is possible to test the Kestrel at humidity values other than 32.8% and 75.3%. To set up a test at another RH value, simply use a different saturated salt solution in the humidity chamber. Refer to the table on page 10 for various salts and their corresponding RH values. To test the accuracy of the Kestrel at various RH values:
|
||
| SALT BATH | PUBLISHED RH AT 25°C |
| LITHIUM BROMIDE | 6.37% |
| LITHIUM CHLORIDE | 11.30% |
| POTASSIUM ACETATE | 22.51% |
| MAGNESIUM CHLORIDE | 32.80% |
| POTASSIUM CARBONATE | 43.16% |
| MAGNESIUM NITRATE | 52.89% |
| SODIUM BROMIDE | 57.57% |
| POTASSIUM IODIDE | 68.86% |
| SODIUM CHLORIDE | 75.30% |
| POTASSIUM CHLORIDE | 84.34% |
| POTASSIUM SULFATE | 97.30% |
*Based on:
Humidity Fixed Points of Binary Saturated Aqueous Solutions,
Greenspan, Lewis; Journal of Research of the National Bureau of Standards-A.
Physics and Chemistry, Vol. 81A, No. 1, January-February 1977
![]()
There are over 400 Pages on this Web site. It is one of the world's largest Renewable Energy web sites. Take some time and check out the amazing and essential products we offer.